212
KHA3000
Operation after Starting the Test (61000-3-3,
Manual SW)
The following figure shows the operation procedures for the case IEC 61000-3-3 (Manual
SW) is selected.
•
The measurement is stopped each time the
“
d Meas Time
”
of one
“
d Meas Count
”
is
completed.
•
During the pause, prepare the EUT for the next operation.
•
According to the dialog and menu displayed, press the appropriate function key to start
the next
“
d Meas Time
”
.
•
After repeating the above, terminate the test when the
“
d Meas Count
”
is completed.
Display of dialog to start
the next measurement
F2 key
Measure the
same segment
F1 key
Measure the
next segment
Next
measurement
Prepare EUT
for operation
START key
d Meas Time
d Meas Count: 1
d Meas Count: 2
d Meas Count: n
Pause
Test complete