192
KHA3000
61000-3-3
Voltage Range, Current Range, Wiring Method and Scaling
No.
Menu item
Description
Default
1
Ch Setting (V/I Range)
Select the input phase for the voltage and current measurement.
This is used when "Independent" is selected in F5 Channel Linkage.
If "Linked" is selected in F5 Channel Linkage, this part is grayed out and
operation is inhibited. When "Linked" is selected, the L1 phase will be
the reference.
L1
F1
→
Sub Menu
F2
→
Sub Menu
F3
→
Sub Menu
F4
F6
F1
L1
L2
L3
F2
150 V
300 V
600 V
AUTO
F3
0.5 A
1 A
2 A
5 A
10 A
20 A
40 A
AUTO
F5
F6
F1
F2
F3
F6
_ Voltage Range
_ Current Range
Linked
Independent
Return
Return
Channel Linkage
Prev/Next
Page 4/4
Wiring Method
1P2W
1P3W
3P3W
3P4W
PLL Source
Voltage L1
Voltage L2
Voltage L3
Current L1
Current L2
Current L3
EXT
Delta Transform
Enable
Disable
1
3
2
4
5
6
7
8
9
10
11
F2
F3
When "Independent" is selected
in Channel Linkage, the input
phase selected (L1,L2 or L3) will
be shown at the underlined position.
_ Voltage Range
_ Current Range
Menu 4/4 (part 1)
V/ I Range Setting
Input/Scaling Setting
Input Config.
DC Offset
Ch Setting
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...