262
KHA3000
In-rush Current Measurement
This function monitors in-rush current waveforms exceeding the trigger level. It can also
monitor voltage waveforms.
After the reset trigger function key is pressed, the present waveform is maintained until the
current value exceeds the trigger level. The scale sensitivity is displayed on the upper part of
the graph according to the selected menu.
Select Trigger Channel, Current Trigger and Reset Trigger/Abort
In-rush Current Measurement sub-menu 1/3
F4
Sub Menu
F1
Select Trigger Ch
L1
L2
L3
F2
Current Trigger (A)
(Input numeric value)
F3
Reset TRG
F3 Abort (Wait for TRG)
F5
Return to Top Menu
F6
Prev/Next
Page 1/3
10,
5,
2,
1,
1/2,
1/5
F2
Vertical Scale (Current)
10,
5,
2,
1,
1/2,
1/4
F3
Vertical Scale (Voltage)
10,
5,
2,
1,
1/2,
1/4
F4
Cursor
(Move the cursor)
F5
Check View Phase
L1
L2
L3
F6
Return
F1
Horizontal Scale
4
5
8
9
6
7
→
View Setting
1
3
3
2
10
11
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......