276
KHA3000
Overview of Remote Control
This product can be operated through the front panel and also remotely through the following
interfaces (standard equipment):
•
RS232C interface
•
GPIB interface
•
USB interface
These interfaces cannot be used concurrently.
The remote interfaces comply with IEEE Std 488.2-1992 and SCPI Specification 1999.0.
Before using SCPI commands, make sure that you understand the SCPI command syntax for
this product.
The RMT LED on the front panel illuminates during operation in remote control mode. To
change remote control mode to local control mode (panel operation) using the front panel,
press the LOCAL key.
Measurement equipment interface standards
This product conforms to the following standards:
•
IEEE Std 488.2-1992 IEEE Standard Codes, Formats, Protocols, and Common
Commands For Use With IEEE Std 488.1-1987
•
IEEE Std 488.1-1987 IEEE Standard Digital Interface for Programmable Instrumentation
•
Standard Commands for Programmable Instruments (SCPI) version 1999.0
•
Universal Serial Bus Specification Rev 2.0
•
Universal Serial Bus Test and Measurement Class Specification (USBTMC) Rev 1.0
•
Universal Serial Bus Test and Measurement Class, Subclass USB488 Specification
(USBTMC-USB488) Rev 1.0
VISA Library
To use a VISA library (VISA COM) as an I/O library, the VISA library must be installed in the
controller (host computer).
When a USB interface is used for control, a device driver compliant with the USB T&M class
(USBTMC) is required. The USBTMC driver is automatically installed by the VISA library.
The Virtual Instrument Software Architecture (VISA) is a standard specification of instrument
connection software established by VXlplug&play Systems Alliance.
One of the VISA libraries (driver software installed according to the VISA specifications) listed
below is required.
•
NI-VISA (ver. 3.2 or later) of National Instruments
•
Keysight VISA (Keysight IO Libraries M01.00 or later) of Keysight Technologies
•
KI-VISA ver. 3.0.0 or later
Do not install multiple VISA libraries in your PC; doing so may cause a malfunction.
See
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......