Report Printing Format
Terms used for the Report Printing Format
This section explains the meanings of terms used for report printing formats. Terms defined in
the related standards are excluded from the explanation.
Harmonics
Related standard
Wiring method
Term
Meaning
Common
Common
Final Test result
General judgment result
PASS, Pass
The result is acceptable.
FAIL, Fail
The result is not acceptable.
WARN, Warn
The result is acceptable but the specified
margin is exceeded.
Order
Harmonic order (2nd to 40th; 1st order is a
fundamental.)
Lim1[Arms]
Limit value
Lim2[Arms]
150 % or 200 % of the limit value
AVE[Arms]
Average value in the measurement time
MAX[Arms]
Maximum value in the measurement time
Judge
Judgment for each harmonic order
Three-phase
L1 Test Result
Judgment result of the L1 input phase
L2 Test Result
Judgment result of the L2 input phase
L3 Test Result
Judgment result of the L3 input phase
Total Current
Total current of all input phases
Total Power
Total power of all input phases
Total Power Factor
Total power factor of all input phases
Total Apparent power
Total apparent power of all input phases
Total Reactive Power
Total reactive power of all input phases
IEC 61000-3-2
Ed3.0
Single-phase,
Three-phase
limitOver[s]
The length of time (seconds) during which
150 % of the limit value is exceeded
IEC 61000-3-12
Single-phase,
Three-phase
Ave[%] Reading(In/Iref)
Ave[%] Reading(In/I1)
Average value in the measurement time and its
ratio to the reference current
Average value in the measurement time and its
ratio to the reference fundamental current
Max[%] Reading(In/Iref)
Max[%] Reading(In/I1)
Maximum value in the measurement time and
its ratio to the reference current
Maximum value in the measurement time and
its ratio to the reference fundamental current
Ave[Arms] Reading
Average value in the measurement time
Max[Arms] Reading
Maximum value in the measurement time
Iequ
Rated current of the EUT
Iref
I1
Reference current
Reference fundamental current
Judgment Rsce
Rsce value used for judgment
Limit values
Limit values to be applied
Ssc
Short-circuit power (calculated value)
Sequ
Rated apparent power of equipment (calculated
value)
Z
System impedance (calculated value)
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......