Report Printing Format
382
KHA3000
IEC 61000-3-3 Ed3.1, single-phase two-wire, Pst Auto
Voltage Fluctuation and Flicker Test Report
N/A : Not Apply
Company
Model Name
Type
Serial No.
Operating Mode
Date of Test
2021/03/11 10:34:45
Climatic Condition
Memo
Test Engineer
Type of Test
EN 61000-3-3(2013)/A1(2019)
*IEC 61000-3-3 Ed3.1(2017)
EN 61000-4-15(2011)
*IEC 61000-4-15 Ed2.0(2010)
Power Analyzer
KHA3000, Ver3.50
Supply Source
Reference Impedance
Wiring Method
Single-Phase 2-Wire
Nominal Voltage
230 V
Nominal Frequency
50 Hz
Pst Measurement Time
30 s
Pst Measurement Count
12
Flicker Margin
100 %
d Measurement Margin
100 %
Test Data of Voltage Fluctuation and Flicker
Final Test Result
PASS
Segment
Pst
dmax[%]
dc[%]
Tmax[ms]
Judge
Limit
1.000
6.000
3.300
500
Seg.1
0.379
0.348
0.186
0
Pass
Seg.2
0.375
0.225
0.031
0
Pass
Seg.3
0.373
0.274
0.197
0
Pass
Seg.4
0.373
0.305
0.134
0
Pass
Seg.5
0.372
0.207
0.090
0
Pass
Seg.6
0.372
0.308
0.203
0
Pass
Seg.7
0.372
0.268
0.063
0
Pass
Seg.8
0.375
0.220
0.162
0
Pass
Seg.9
0.372
0.363
0.191
0
Pass
Seg.10
0.377
0.251
0.005
0
Pass
Seg.11
0.372
0.266
0.185
0
Pass
Seg.12
0.373
0.329
0.157
0
Pass
Plt
Value
Judge
Limit
0.650
Measurement
0.374
Pass
Page 1/1
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......