28
KHA3000
Three-phase Four-wire Circuit
Wiring Method
Use the L1, L2, L3, N, and G terminals in the input section on the rear of this product.
Connect them according to the following table.
If LIN is not used, connect the wires directly from the output terminals of the AC power supply
to the SOURCE terminals in the input section of this product.
A
V
A
V
A
V
U
V
W
N
G
U
V
W
N
G
N
G
L1
L2
L3
KHA3000
SOURCE
LOAD
EUT
AC power
supply
Impedance
network
CAUTION
Terminal G of this product is not a protective conductor terminal. It is not grounded even if
the wire is connected to terminal G. For terminals G and N, the respective SOURCE and
LOAD terminals are short-circuited internally.
AC power
–
LIN
–
Input section on
the rear of KHA3000
–
EUT terminal
Output terminal
INPUT
OUTPUT
SOURCE
LOAD
U
–
U
U
–
L1
L1
–
U
V
–
V
V
–
L2
L2
–
V
W
–
W
W
–
L3
L3
–
W
N
–
N
N
–
N
N
–
N
G
–
G
G
–
G
G
–
G
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......