5
KHA3000
111
Setting the Harm
onic Test
61000-3-12
Limitation Standards, Equipment and Rated of Equipment
No.
Menu item
Description
Default
1
Limitation Std
Select limit value standard IEC 61000-3-12 2004 or IEC 61000-3-12
2011. If a standard other than IEC 61000-3-12 2004 and IEC 61000-3-12
2011 is selected, the menu for equipment with input current 16, 20 A or
less is selected. Abbreviated Standard names are shown on the display.
IEC 61000-3-2
Ed5.0
2
Meas Technic
Select a measuring technique standard. IEC 61000-4-7 Ed2.1 and IEC
61000-4-7 Ed2.0 measure interharmonic waves and specifies the result
as a harmonic group value. IEC 61000-4-7 Ed2.1 or IEC 61000-4-7
Ed2.0 are recommended for measuring fluctuating harmonics.
Abbreviated Standard names are shown on the display.
IEC 61000-4-7
Ed2.1
3
Equipment
Select the type of EUT.
Single-phase equipment includes the single-phase part of hybrid
equipment. Balanced three-phase equipment includes the three-phase
part of hybrid equipment. This item is used for calculating the Sequ value.
Single-phase
equipment
4
Rated Voltage (V)
Enter the rated voltage of the EUT.
For the rated voltages of single-phase and unbalanced three-phase
equipment, enter phase voltages. For the rated voltages of three-wire/
three-phase equipment and balanced three-phase equipment, enter
line voltages. This item is used for calculating the Sequ value.
The input range is 100 V to 600 V.
Enter the value with the numeric keypad and press the ENTER key, or
use the large knob to increase or decrease the value.
230 V
5
Rated Current (lequ)
Specify the rated current (lequ). Select
“
Measured
”
or
“
Specified.
”
Measured value
F1
Limitation Std
IEC 61000-3-2 Ed3.0
IEC 61000-3-12 2011
IEC 61000-3-2 Ed5.0
IEC 61000-3-2 Ed2.2
JIS 61000-3-2 2005
JIS 61000-3-12 2019
IEC 61000-3-12 2004
F2
Meas Technic
IEC 61000-4-7 Ed2.0
IEC 61000-4-7 Ed2.1
IEC 61000-4-7 Ed1.0
F3
Equipment
Single-Phase
3-Wire, 3-Phase
Balanced 3P
Unbalanced 3P
F4
Rated of Equipment
Sub Menu
F5
F6
F1
Rated Voltage (V)
(Input numeric value)
F2
Rated Current (lequ)
Measured value
Specified value
F3
Rated Current lequ (A)
(Input numeric value)
F4
Nominal Sys Volt
(V)
Unom
(Input numeric value)
F5
Nominal Frequency
50 Hz
60 Hz
F6
Return
To Meas Setting Page
→
1
3
2
4
5
7
6
8
9
10
11
Menu 1/4
Prev/Next
Page 1/4
IEC 3-2 Ed3.0
IEC 3-12 2011
IEC 3-2 Ed5.0
IEC 3-2 Ed2.2
JIS 3-2 2005
IEC 3-12 2019
IEC 3-12 2004
Display of abbreviated
Limitation Standard names
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......