15
KHA3000
355
Specifications
Specifications
Measurement and Display Items
Voltage Measurement Function
Item
Specifications
Voltage
Real effective value and peak value (+ peak and - peak)
Current
Real effective value and peak value (+ peak and - peak)
display
Voltage
Vrms
Single-phase three-wire: (L1 Vrms +L2 Vrms)/2
Three-phase three-wire,Three-phase four-wire: (L1 Vrms +L2 Vrms
+L3 Vrms) /3
Current
Arms
Single-phase three-wire: (L1 A rms +L2 Arms)/2
Three-phase three-wire,Three-phase four-wire: (L1 Arms +L2 Arms
+L3 Arms) /3
Apparent power
VA
Single-phase three-wire: L1 VA +L2 VA
Three-phase three-wire:
(L1 VA +L2 VA +L3 VA) × (
/3)
Three-phase four-wire: (L1 VA +L2 VA +L3 VA)/3
Real power
W
Single-phase three-wire: L1 W +L2 W
Three-phase three-wire: L1 W +L3 W
Three-phase four-wire: L1 W +L2 W +L3 W
Reactive power
var
Single-phase three-wire: L1 var +L2 var
Three-phase three-wire: L1 var +L3 var
Three-phase four-wire: L1 var +L2 var +L3 var
THD (total harmonic distortion)
Effective value ratio of all the harmonic components from the 2nd to
the 40th order against the effective value of the fundamental
components current
THDI (total harmonic distortion)
Effective value ratio of all the harmonic components from the 2nd to
the 40th order against the effective value of the current
PWHD (partial-weighted harmonic
distortion)
Effective value ratio of all the harmonic components from the 14th to
40th order, which are weighted according to the order, against the
effective value of the fundamental components
Item
Specifications
Range
150 V/300 V/600 V
Allowable crest
factor
150 V and 300 V
range
2
600 V range
2
*1
*1.
Up to
±
900 Vpeak
Display item
TrueRMS and
±
peak
Accuracy
*2
*2.
Accuracy is not specified for
±
Peak. PLL is in the Lock state, in the effective input range, and with anti-
aliasing filter cutoff frequency 6 kHz. Specified for the ambient temperature of 18
°C
to 28
°C
and relative
humidity of 20 % to 80 % (without condensation). When scaling is used, the voltage range shown on the
display is the "range" value.
±
(0.4 % of rdng+0.04 % of range)
Resolution
(RMS/Peak)
150 V range
0.01 V/0.01 V
300 V range
0.01 V/0.02 V
600 V range
0.01 V/0.04 V
Display range
(RMS/Peak)
150 V range
120 % of the range/213 % of the range
300 V range
120 % of the range/213 % of the range
600 V range
105 % of the range/151 % of the range
Effective input range
60 % to 110 % of the range
Excess input display
“
Over
”
is displayed when the
±
peak value or RMS value exceeds
the display range.
Fundamental frequency range
45 Hz to 65 Hz
3
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......