FUNC
Sets the operation mode.
Command
[SENSe:]FUNCtion[:ON]
“{HARMonics|VFL|OTHer:BASic|OTHer:FFT|OTHer:RUSH}”
[SENSe:]FUNCtion[:ON]?
Parameter
“HARMonics”
Harmonic measurement mode (default)
“VFL”
Voltage fluctuation measurement mode
“OTHer:BASic”
Basic measurement
“OTHer:FFT”
FFT analyzer
“OTHer:RUSH”
In-rush current measurement
Response
Returns the present operation mode in the <string> format.
* RST
HARM:MTIM
Sets the measurement time.
Command
HARMonics:MTIMe {<numeric>|MIN|MAX}
HARMonics:MTIMe?
Parameter
1 to 9600 (The default is 150.)
Unit:
S
Response
Returns the measurement time in the <NR3> format.
* RST
Value:
Value:
Attach a value you want to set after a
command and send the command. To set
the measurement time at 20 seconds,
send HARM:HTIM 20.
The set value of a command provided with these marks is
influenced by *RST or *RCL when it is sent and changed to
the value described in “Default Conditions.”
The test time is changed to 150 seconds when *RST is sent.
A command is described in the long
form. The lower-case part can be
omitted. Parts enclosed in [ ] can
also be omitted.
This is the range of values
that can be set.
Optional symbols such as
m and μ can be used.
This is the unit of the
value to be set. The unit
can be omitted.
If MIN is specified, the minimum value is applied.
If MAX is specified, the maximum value is applied.
This is the expression format of the
value when returned in response to
the query that has been sent.
This is the expression format of the value when
returned in response to a query that has been sent.
If a parameter is too long to
fit on one line, a new line is
inserted.
Parameters are described. In the case
of this example, because the parameter
is “numeric,” a specific value can be
specified. The minimum or maximum
value can also be specified.
Section
Page
Command syntax
Parameter
Unit
Default conditions
Query
Messages
Error List
See
** RST
RST
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......