Setting the Harm
onic Test
●
Observation period (Tobs)
“
Tobs
”
is the observation period of a test that is specified by a standard. A period of time
required by the operation type of the EUT is specified as the observation period. There
are four categories of observation type:
“
quasi-static,
”
“
short cycle,
”
“
random
”
and
“
long
cycle,
”
according to the type of harmonic current fluctuation.
To specify the test observation period, check the operation type of the EUT and set a
measurement time that is longer than the observation period required for the operation type.
●
Standard names and their standard presentations
Procedures for inputting a character (F1 key) and deleting a character (F2 key)
1
Press the F1, F2, F3 or F4 key.
The Comment Input dialog will be presented.
2
Use the small knob, large knob or arrow keys to select a character by
moving the rectangular frame.
3
Press the F1 (Input 1-char) key.
The character selected will be added to the part where the cursor is blinking (Input 1-
char). To delete a character, press the F2 (Delete 1-char) key. To change the character
type, press the F3 (Change char type) key to select the type. To repeat adding or
deleting characters, return to the above step 2.
4
Press the ENTER key.
The characters entered will be fixed and saved.
8
Return
Terminates the comment input.
–
9
Std. Name Select
Select how the "Limitation Std" and "Meas. Tech" names are
presented in Report Print. Selects either "Standard" or "Setting".
Standard
10
Input 1-char
This function is used to enter characters for the local Limitation Standard
name or the local Measuring technique name to be presented in Report
Print. Up to 20 alphanumeric characters can be entered. On how to use
this function, refer to the procedures shown below.
–
11
Delete 1-char
This function is used to delete characters from the local Limitation
Standard names or the local Measuring technique names to be
presented in Report Print. On how to use this function, refer to the
procedures shown below.
–
12
Return
Exit from the character entry/deletion procedures to display local
Limitation Standard or local Measuring technique names.
–
13
Return
Exit from the setting of standard names to be presented.
–
14
Reference Impedance
Set whether to use the reference impedance.
Unuse
15
Prev/Next Page 2/4
Switches to another menu page.
–
F key: Function key. You can also use the ESC key to return.
Menu presentation
Standard presentation for Report Print
*1
*1.
At factory shipment, the default names presented for the "Setting" mode
are the same as Standard names.
IEC 61000-3-2 Ed3.0
EN 61000-3-2(2006)
IEC 61000-3-2 Ed5.0
EN IEC 61000-3-2(2019)
IEC 61000-3-2 Ed2.2
EN 61000-3-2(2000)/A2(2005)
JIS 61000-3-2 2005
JIS C61000-3-2(2005)
JIS 61000-3-2 2019
JIS C61000-3-2(2019)
IEC 61000-3-12 2004
EN 61000-3-12(2005)
IEC 61000-3-12 2011
EN 61000-3-12(2011)
IEC 61000-4-7 Ed2.0
EN 61000-4-7(2002)
IEC 61000-4-7 Ed2.1
EN 61000-4-7(2002)/A1(2009)
IEC 61000-4-7 Ed1.0
EN 61000-4-7(1993)
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......