5 Installation
35
Flowmeter sensor
1
Terminals for signal cable
2
Terminal for equipotential bonding
Figure 43: Sensor connection for remote mount design (example)
Change from one to two columns
NOTICE
If the O-ring gasket is seated incorrectly or damaged, this
may have an adverse effect on the housing protection
class.
Follow the instructions in
Opening and closing the terminal
box
on page 22 to open and close the housing safely.
Compliance with the information below is required for making the
electrical connections:
• Route the signal cable into the housing as shown.
• Connect the cables in accordance with the connection
diagrams. Connect the cable shields (if applicable) to the
provided grounding clamp.
• Use ferrules for making the connections.
• Use suitable plug(s) to seal the cable entries that are not
used.