5 Installation
19
5.1.10 Free inlet or outlet
Figure 16: Free inflow and outflow
A
For a free outflow, do not install flowmeter at the highest point
of the piping or on its outflow side, since the measuring tube
may run empty, creating air bubbles.
B
For free inflow/outflow, provide an invert to make sure that the
piping is always full
5.1.11 Mounting with heavily contaminated
measuring media
Figure 17: Bypass line
For strongly contaminated measuring media, a bypass line in
accordance with the figure is recommended so that operation of
the system can continue to run without interruption during
mechanical cleaning.
5.1.12 Mounting with pipe vibration
1
Pump
2
Damping device
3
Turn-off device
Figure 18: Vibration damping
If pipe vibration occurs, it needs to be damped using damping
devices.
The damping devices must be installed outside the support
section and outside of the piping section between the turn-off
devices.
Avoid connecting damping devices directly to the flowmeter
sensor.
Summary of Contents for 406060
Page 2: ......
Page 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Page 99: ...13 Appendix 99...
Page 100: ...13 Appendix 100...
Page 101: ...13 Appendix 101...
Page 102: ...13 Appendix 102...
Page 103: ...13 Appendix 103...
Page 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Page 105: ...13 Appendix 105...
Page 106: ...13 Appendix 106...
Page 107: ...13 Appendix 107...
Page 108: ...13 Appendix 108...
Page 109: ...13 Appendix 109...
Page 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Page 111: ...13 Appendix 111...
Page 112: ...13 Appendix 112...
Page 113: ...13 Appendix 113...
Page 115: ......