6 Commissioning
36
6 Commissioning
6.1 Safety
instructions
CAUTION
Risk of burns due to hot measuring media
The device surface temperature may exceed 70 °C (158 °F),
depending on the measuring medium temperature!
• Before starting work on the device, make sure that it has
cooled sufficiently.
Aggressive or corrosive media may lead to the damage of wetted
parts of the sensor. As a result, measuring medium under
pressure can leak out.
Wear to the flange gasket or process connection gaskets
(e.g. pipe fitting, Tri-clamp, etc.) may caused a pressurized
measuring medium to escape.
When using internal flat gaskets, they can become brittle through
CIP- / SIP processes.
If pressure surges above the permissible nominal pressure of the
device occur permanently during operation, this may affect the
service life of the device.
If there is a chance that safe operation is no longer possible, take
the device out of operation and secure it against unintended
startup.
6.2
Write-protection switch, service LED and
local operator interface
1
Write-protection switch
2
Service LED
3
Local operator interface
Figure 44: Write-protection switch, service LED, local operator interface
Write-protection switch
With write protection activated, device parameter settings cannot
be modified via the local operator interface or the local display.
Turn the write-protection switch clockwise to deactivate write
protection and turn it counterclockwise to activate write
protection.
Service LED
The service LED that indicates the operating state of the device is
located in the terminal box for the sensor.
Service LED
Description
Flashing rapidly (100 ms)
Start-up procedure in progress; device is not ready
for operation
Permanently lit
Device is running; no critical error
Flashing slowly (1 second) No critical error has occurred; see
Diagnosis /
error messages
on page 68
Summary of Contents for 406060
Page 2: ......
Page 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Page 99: ...13 Appendix 99...
Page 100: ...13 Appendix 100...
Page 101: ...13 Appendix 101...
Page 102: ...13 Appendix 102...
Page 103: ...13 Appendix 103...
Page 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Page 105: ...13 Appendix 105...
Page 106: ...13 Appendix 106...
Page 107: ...13 Appendix 107...
Page 108: ...13 Appendix 108...
Page 109: ...13 Appendix 109...
Page 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Page 111: ...13 Appendix 111...
Page 112: ...13 Appendix 112...
Page 113: ...13 Appendix 113...
Page 115: ......