8 Diagnosis / error messages
69
8.2 Error
messages
The error messages are divided into four groups in accordance with the NAMUR classification scheme.
Errors
Error no. / Area
Text on the LCD display
Cause
Remedy
F099.042 / Electronics
ADC overrange alarm.
Noise too high. Check applicati.
Call service.
The signal at the input of the A / D converter
exceeds the maximum value.
Flow measurement is not possible.
When the meter tube is empty, ensure that the
empty pipe detection function is active.
Ensure that the current flow rate does not exceed
the set measuring range.
Contact the manufacturer.
F098.012 / Electronics
FEB communication error.
EMC disturbance. Call Service.
Defective front end board or motherboard.
Wiring error between sensor and transmitter.
The front end board and the motherboard are not
connected.
If this involves an integral mount design: Contact
the manufacturer.
If this involves a remote mount design: Check the
wiring between the sensor and the transmitter.
Contact the manufacturer.
F097.029 / Electronics
Coil regulation error.
Check wiring of sensor coils. Call
service
Defective solenoid.
Wiring error, broken cable or shorted out coil circuit
(terminals M1 / M2).
Defective fuse for coil circuit; moisture in sensor
terminal box.
Check wiring.
Check coil circuit fuse.
Check the sensor terminal box for moisture.
F096.043 / Electronics
SIL self check alarm.
Call service.
The SIL monitoring function has detected a
transmitter error.
Contact the manufacturer.
F095.036 / Electronics
Coil isolation alarm.
Call service.
Defective solenoid.
Wiring error with coil circuit (short-circuit, terminals
M1 / M2 downstream of GND).
Flooded sensor.
Contact the manufacturer.
F094.022 / Electronics
Safety Alarm Curr. Out 31 / 32
SIL function detects error. Call
Service.
SIL-relevant error detected at current output.
Contact the manufacturer.
F093.032 / Electronics
Electrode short cuircit.
Check wiring of sensor electrode.
Call service.
Wiring error with electrode circuit (short-circuit
downstream of shield).
Leaking electrode.
Flooded sensor.
Check wiring.
Check electrode wiring for short-circuit to shield.
Check electrode for leakage.
Check the sensor terminal box for moisture.
F092.033 / Electronics
Electrode open cuircit.
Check wiring of sensor electrode.
Call service.
Wiring error with electrode circuit (interruption,
broken cable).
Check wiring.
F091.030 / Electronics
Coil wiring error.
Check wiring of sensor coils. Call
service.
Defective solenoid.
Wiring error, broken cable or shorted out coil circuit
(terminals M1 / M2).
Defective fuse for coil circuit; moisture in sensor
terminal box.
Check wiring.
Check coil circuit fuse.
Check the sensor terminal box for moisture.
F090.035 / Electronics
ADC RX210 com. error.
Call service.
Problems with the electromagnetic compatibility
(EMC) in the vicinity.
Error with transmitter electronics.
Replace the transmitter or contact the
manufacturer.
F088.013 / Electronics
FEB communication error.
EMC disturbance. Call Service.
Problems with the electromagnetic compatibility
(EMC) of the signal cable.
Incorrect signal cable.
Check the signal cable and the wiring.
Contact the manufacturer.
Summary of Contents for 406060
Page 2: ......
Page 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Page 99: ...13 Appendix 99...
Page 100: ...13 Appendix 100...
Page 101: ...13 Appendix 101...
Page 102: ...13 Appendix 102...
Page 103: ...13 Appendix 103...
Page 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Page 105: ...13 Appendix 105...
Page 106: ...13 Appendix 106...
Page 107: ...13 Appendix 107...
Page 108: ...13 Appendix 108...
Page 109: ...13 Appendix 109...
Page 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Page 111: ...13 Appendix 111...
Page 112: ...13 Appendix 112...
Page 113: ...13 Appendix 113...
Page 115: ......