7 Operation
59
Menu / Parameters
Description
Device Setup / ...Transmitter / ...Units
Unit Massflow Qm
Used to select the unit for the mass flow.
Refer to
Available units
on page 53.
The selection applies to the current mass flow display and to the parameters that relate to the mass flow such as QmMax and
Qm
Max
DN.
Unit Mass Tot.
Used to select the unit for the mass meter.
Refer to
Table 3: Units for the mass totalizer
on page 53.
Unit Volumeflow Qv
Used to select the unit for the volumetric flow.
Refer to
Table 1: Units for the volume flow
on page 53.
The selection applies to the current volumetric flow display and to the parameters that relate to the volumetric flow rate such
as QvMax and QvMaxDN.
Unit Vol. Totalizer
Used to select the unit for the volumetric meter.
Refer to
Table 4: Units for the volume totalizer
on page 53.
Cust Qm Tot Unit F.
Used to enter the factor for a user-defined unit of mass.
This factor relates to the flow per liter.
Cust Qm Tot Unit N.
Used to enter the name of the user-defined unit of mass.
Cust Qv Tot Unit F.
Used to enter the factor for a user-defined unit of volume.
This factor relates to the flow per liter.
Cust Qv Unit N.
Used to enter the name of the user-defined unit of volume.
Device Setup / ...Transmitter / ...Cut Off
Low Flow Cut Off
Used to set the switching threshold (0 to 10 %) for low flow suppression.
If the flow rate is below the set switching threshold for low flow, the flow is not measured. The setting of 0 % deactivates the
low flow suppression option.
Default setting: 1.0 %
Low Flow Hysteresis
Used to set the hysteresis (0 to 50 %) for low flow suppression as defined by the “Low Flow Cut Off” parameter setting.
Default setting: 20 %
Device Setup / ...Transmitter / ...Feature Settings
Verifying Capab.
Used to indicate whether the verification function is active.
Verifying Capab.Code
Used to set the device-specific code to activate the verification function. To retrofit this option, contact the manufacturer or the
sales organization.
Restart the device after the code has been entered (e.g. via the “Device Reset” parameter or by briefly switching off the
power supply).
Device Setup / ...System Zero
Manual Adjust
Used to set the value for zero point calibration in % of Q
max
DN.
Manual setting: -50 to +50 mm/s
Auto Adjust
Used to start the automatic zero point calibration process with
. The automatic zero point calibration process takes
approx. 60 seconds.
Note
Prior to starting the zero point calibration process, make sure that:
• There is no flow through the sensor (close all valves, shut-off devices, etc.).
• The sensor is filled completely with the medium being measured.
Summary of Contents for 406060
Page 2: ......
Page 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Page 99: ...13 Appendix 99...
Page 100: ...13 Appendix 100...
Page 101: ...13 Appendix 101...
Page 102: ...13 Appendix 102...
Page 103: ...13 Appendix 103...
Page 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Page 105: ...13 Appendix 105...
Page 106: ...13 Appendix 106...
Page 107: ...13 Appendix 107...
Page 108: ...13 Appendix 108...
Page 109: ...13 Appendix 109...
Page 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Page 111: ...13 Appendix 111...
Page 112: ...13 Appendix 112...
Page 113: ...13 Appendix 113...
Page 115: ......