2 Design and function
10
2.2 Model
variants
Integral mount design
The transmitter and the flow sensor form a mechanical unit in
devices featuring a compact (integral mount) design.
Remote mount design
The transmitter and the flow sensor are mounted physically
separated from one another in devices featuring a separate
(remote mount) design.
The electrical connection between the transmitter and the sensor
is provided by a signal cable.
With a minimum conductivity of 20 µS/cm of the medium being
measured, a maximum signal cable length of 50 m (164 ft) is
possible.
Information regarding JUMO flowTRANS MAG S10
JUMO flowTRANS MAG S10 flow sensors are available in two
different versions, variant 1 and variant 2. See
Design and
function
on page 7.
2.3 Measuring
principle
Measurements performed by the electromagnetic flowmeter are
based on Faraday’s law of induction. A voltage is generated in a
conductor when it moves through a magnetic field.
1
Magnet coil
2
Measuring tube in electrode plane
3
Measuring electrode
Figure 3: Electromagnetic flowmeter diagram
v
D
B
U
~
1
v
D
qv
4
2
qv
U ~
1
U
1
Measuring span
B Magnetic
induction
D Electrode
spacing
v Average flow velocity
qv Volume flow rate
With the device-relevant application of this measuring principle, a
conductive measuring medium flows through a tube in which a
magnetic field is generated perpendicular to the flow direction
(see Figure 3).
The voltage induced in the measuring medium is tapped by two
diametrically opposed electrodes. This measurement voltage is
proportional to the magnetic induction, the electrode spacing and
the average medium velocity v.
Taking into account that the magnetic induction and the electrode
spacing are constant values results in a proportion between the
measurement voltage U
1
and the average medium velocity.
From the calculation of the volume flow rate follows that the
measurement voltage is linear and proportional to the volume
flow rate
The induced voltage is converted by the transmitter to
standardized, analog and digital signals.
Summary of Contents for 406060
Page 2: ......
Page 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Page 99: ...13 Appendix 99...
Page 100: ...13 Appendix 100...
Page 101: ...13 Appendix 101...
Page 102: ...13 Appendix 102...
Page 103: ...13 Appendix 103...
Page 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Page 105: ...13 Appendix 105...
Page 106: ...13 Appendix 106...
Page 107: ...13 Appendix 107...
Page 108: ...13 Appendix 108...
Page 109: ...13 Appendix 109...
Page 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Page 111: ...13 Appendix 111...
Page 112: ...13 Appendix 112...
Page 113: ...13 Appendix 113...
Page 115: ......