8 Diagnosis / error messages
71
Error no. / Area
Text on the LCD display
Cause
Remedy
S046.001 / Operation
Mass flowrate exceeds limits.
Check flowrate and alarm limits.
The mass flow is above or below the configured
limit values for “Qm Massflow Min” and “Qm
Massflow Max”.
Check the parameter settings in the “Process
Alarm / ...Alarm Limits” menu and adjust as
needed.
Check the mass flow.
S044.002 / Operation
Volume flowrate exceeds limits.
Check flowrate and alarm limits.
The mass flow is above or below the configured
limit values for “Qv Volumeflow Min” and “Qv
Volumeflow Max”.
Check the parameter settings in the “Process
Alarm / ...Alarm Limits” menu and adjust as
needed.
Check the volumetric flow.
S041.034 / Electronics
DC feedback regulation.
Check conditions of application.
Call service.
Multi-phase media are being measured that
produce a very high level of noise.
Stones or solids that produce a very high level of
noise.
Galvanic voltages at the measuring electrodes.
Conductivity of medium being measured is not
evenly distributed (e.g., directly after injection
points).
Contact the manufacturer.
S040.031 / Electronics
Coil impedance alarm.
Call service.
Change in coil inductance; damaged coil;
damaged coil insulation; interference due to
external magnetic fields.
Contact the manufacturer.
Maintenance required
Error no. / Area
Text on the LCD display
Cause
Remedy
M038.010 / Electronics
Sensor memory defective.
Mem. or connect. defective.
Replace memory.
No SensorMemory.
Defective SensorMemory.
Check that the correct SensorMemory is inserted.
Contact the manufacturer.
M037.015 / Electronics
NV chips defect on Motherboard.
Defective MB. Replace MB. Call
Service.
Defective SensorMemory in motherboard
Replace the transmitter.
Contact the manufacturer.
M032.023 / Config.
Curr.Out 31 / 32 not calibrated.
Call Service.
Current output 31 / 32 is not calibrated.
Contact the manufacturer.
M028.008 / Config.
Display value is < 1600 h at
Qmax.
Change mass Unit or vol. Unit for
Totalizer.
The current meter reading exceeds the display
resolution
Check the settings for the mass / volumetric meter
and adjust as needed
M026.005 / Config.
Maintenance interval is reached.
Perform maintenance.
Maintenance interval reached.
Set “Preset Maint. cycle” to zero to deactivate the
maintenance counter.
Complete the maintenance tasks.
Start new maintenance interval.
M024.009 / Config.
Device not calibrated.
Call Service.
Device is not calibrated.
Contact the manufacturer.
Summary of Contents for 406060
Page 2: ......
Page 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Page 99: ...13 Appendix 99...
Page 100: ...13 Appendix 100...
Page 101: ...13 Appendix 101...
Page 102: ...13 Appendix 102...
Page 103: ...13 Appendix 103...
Page 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Page 105: ...13 Appendix 105...
Page 106: ...13 Appendix 106...
Page 107: ...13 Appendix 107...
Page 108: ...13 Appendix 108...
Page 109: ...13 Appendix 109...
Page 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Page 111: ...13 Appendix 111...
Page 112: ...13 Appendix 112...
Page 113: ...13 Appendix 113...
Page 115: ......