JD780A Series Analyzers
Using Channel Scanner
541
Procedure
To automatically set the reference and attenuation level:
1.
Press the
AMP/SCALE
hard key.
2.
Press the
Auto Scale
soft key.
Each time you press this key, both of the Y-axis scale and input attenuation level change to be
optimized with some margin.
To set the reference or attenuation level manually:
1.
Press the
AMP/SCALE
hard key.
2.
To set the maximum reference value on the Y-axis manually, complete the following steps:
a.
Press the
Reference Level
soft key.
b.
Enter a value by using the numeric keys or the rotary knob with 10 dB increments.
c.
Press the unit soft key or the
ENTER
hard key.
3.
To set the attenuation option, select one from the following choices:
–
To set the input attenuator
’s level automatically, select
Attenuation > Auto
.
NOTE
It is recommended that you set the
Attenuation
to
Auto
in most situations so that the
level of the input attenuator can be set automatically according to your input signal level.
–
To set the input attenuation manually up to 55 dB to optimize S/N, complete the following
steps:
a.
Select
Attenuation > Manual
.
b.
Press the
Attenuation Value
soft key to set the level.
c.
Enter a value between 0 and 55 in fives by using the numeric keys.
d.
Press the
dB
soft key or the
ENTER
hard key.
–
To couple the input attenuator
’s level with your reference level setting, select
Attenuation >
Couple
.
As you increase the reference setting, the attenuation level also increases acoordingly.
Optional
. To change the scale unit:
1.
Select
More (1/2) > Units
.
2.
Select the unit of the display scale:
dBm
,
dBV
,
dBmV
,
dBµV
,
V
, or
W
.
The scale unit on the screen changes accordingly.
Pre-amplifier
You can turn the internal pre-amplifier on to correct and compensate for the gain of the preamp so that
amplitude readings show the value at the input connector.
Procedure
1.
Press the
AMP/SCALE
hard key.
2.
Toggle the
Preamp
soft key and select
On
or
Off
as needed.
NOTE
You can turn the Preamp on when the input attenuation range is from 0 dB to 10 dB. If
the attenuation value is manually set to greater than 10 dB, the instrument will
automatically turn off the pre-amplifier to display low-level signal properly on the chart.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...