JD780A Series Analyzers
Using LTE
– TDD Signal Analyzer
283
2.
To set the nominal channel bandwidth to be measured, complete the following steps:
a.
Press the
Bandwidth
soft key.
b.
Select the bandwidth option:
1.4 MHz
,
3 MHz
,
5 MHz
,
10 MHz
,
15 MHz
, or
20 MHz
.
3.
To select the subframe number to be measured, complete the following steps:
a.
Press the
Subframe No
soft key.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Press the
Enter
soft key.
4.
Toggle the
Cell ID
soft key and select
Auto
or
Manual
as desired:
–
Auto
: to let the instrument detect the Cell ID for the LTE signal automatically.
–
Manual
: to set a specific Cell ID for the LTE signal manually in order to speed up the
synchronization with a BTS.
5.
Optional
. Press the
Miscellaneous
soft key, and then do the following(s) as needed:
–
To select the number of antenna ports, toggle the
MIMO
soft key and select
2x2
or
4x4
.
This MIMO (Multiple Input Multiple Output) setting is activated only if the option 031 LTE-
Advanced TDD is installed with a license number. If not, the instrument sets this option to
2x2 by default. A 2x1 or 4x1 RF combiner is also required to able to test on MIMO
channels.
–
To assign a antenna port number automatically or manually, press the
Antenna Port
soft
key and select the option:
Auto
,
0
,
1
,
2
, and
3
. If the option 031 LTE-Advanced TDD is not
installed or if the MIMO is set to 2x2, the antenna ports 2 and 3 are disabled.
–
To select the cyclic prefix, toggle the
Cyclic Prefix
soft key and select
Normal
or
Extended
. See
“Appendix E – Cyclic prefix table” on page 575 for OFDM parameter details
by different bandwidth.
–
To set the delay, complete the following steps:
a.
Press the
Delay
soft key to set the amount of delay in µs.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Press the
µs
soft key.
NOTE
The Delay setting is used only when there is a time offset in the signals to be measured.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...