![JDS Uniphase CellAdvisor JD780A Series User Manual Download Page 306](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616306.webp)
JD780A Series Analyzers
306
Using LTE
– TDD Signal Analyzer
Signal performance test
Unlike the other auto measurement tests, the Signal Performance Test runs measurements
continuously.
Setting limit for signal performance test
You can set test limits for test item(s) in the auto measurement.
Procedure
1.
Press the
LIMIT
hot key.
2.
Press the
RF Test Limits > Channel Power
soft key, and then set the limits for the channel
power. Other RF test items are not used in the signal performance test.
3.
Press the
Modulation Test Limits
soft key, and then enable test limits as desired.
4.
Optional
. You can enable alarm sound that goes off if the measurement falls outside of the limit.
Toggle the
Beep
soft key between
On
and
Off
to enable or disable the beep sound.
5.
Optional
. Go to
SAVE/LOAD
> Save
, and then select
Limit
to save the limit settings.
See
“Using save” on page 31 for more information.
Setting measure setup
After configuring test parameters as described in the
“Configuring test parameters” on page 265, you
can continue your measurement. The measurement settings can be saved and recalled as a file. You
can also use JDViewer, PC application software to configure a measure setup, save as a file, and load
the file on to the instrument.
Procedure
1.
Press the
MEASURE SETUP
hot key.
2.
Toggle the
Cell ID
soft key and select
Auto
or
Manual
as desired:
–
Auto
: to let the instrument detect the Cell ID for the LTE signal automatically.
–
Manual
: to set a specific Cell ID for the LTE signal manually in order to speed up the
synchronization with a BTS.
3.
Press the
MIMO
soft key, and then select the antenna number option:
2x2
or
4x4
.
This MIMO (Multiple Input Multiple Output) setting is activated only if the option 031 LTE-
Advanced TDD is installed with a license number. If not, the instrument sets this option to 2x2 by
default. A 2x1 or 4x1 RF combiner is also required to able to test on MIMO channels.
4.
Press the
Antenna Port
soft key, and then select the option:
Auto
,
0
,
1
,
2
, or
3
.
If the option 031 LTE-Advanced TDD is not installed or if the MIMO is set to 2x2, the antenna
ports 2 and 3 are disabled.
5.
Press the
Bandwidth
soft key, and then select the nominal channel bandwidth to be measured,
from the following choices:
–
1.4 MHz
,
3 MHz
,
5 MHz
,
10 MHz
,
15 MHz
, and
20 MHz
6.
Toggle the
RF Profile
soft key between
On
and
Off
to turn on or off the RF profiling such as
channel power and subframe frequency error.
7.
Toggle the
Modulation Profile
soft key between
On
and
Off
to turn on or off the modulation
profiling such as P-SS, S-SS, Data - QPSK, Data
– 16 QAM, and Data – 64 QAM.
8.
Toggle the
MIMO Profile
soft key between
On
and
Off
to turn on or off the MIMO profiling such
as TX1 - RS0, TX2 - RS1, and time alignment error.
9.
Toggle the
External Offset
soft key and select
On
to set the external offset value.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...