![JDS Uniphase CellAdvisor JD780A Series User Manual Download Page 102](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616102.webp)
JD780A Series Analyzers
102
Using Interference Analyzer
a.
Toggle the
Period
soft key and select
Manual
.
b.
Press the
Period
soft key to enter the amount of time.
c.
Enter a value by using the numeric keys.
d.
Select the unit:
Sec
,
mSec
, or
µSec
.
5.
Press the
TRIGGER
hot key, and then do one of the following:
–
Select
External
for an external trigger source.
–
Select
Video
or
Display
Position
for the internal trigger source.
6.
Press the
PREV
hard key to continue setting the gated sweep.
7.
Set the gate delay by completing the following steps:
a.
Press the
Gate Delay
soft key.
b.
Enter a value by using the numeric keys.
c.
Select the unit:
Sec
,
mSec
, or
µSec
.
8.
Set the gate length by completing the following steps:
a.
Press the
Gate Length
soft key.
b.
Enter a value by using the numeric keys.
c.
Select the unit:
Sec
,
mSec
, or
µSec
.
9.
Press the
PREV
hard key to enable gated sweeping.
10.
Toggle the
Gated Sweep
soft key and select
On
.
NOTE
This setting is not used in the Spectrum Replayer mode.
Conducting spectrum measurements
The spectrum measurement with an audible indicator is especially useful for locating interferer sources
with a directional antenna.
Setting trace
You can display up to six traces on the measurement chart simultaneously.
Procedure
1.
Press the
TRACE/DISPLAY
hard key.
2.
Press the
Select Trace
soft key, and then select the trace number:
T1
,
T2
,
T3
,
T4
,
T5
, or
T6
.
The legend shape of the selected trace changes from square to round to indicate that the trace is
the active one now.
3.
Do one of the following:
To
Select
Trace Legend
Clear current data and display with new
measurements
Clear Write
W
Display the input signal
’s maximum response only
(unlimited or for a certain amount of time)
Max Hold
M
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...