![JDS Uniphase CellAdvisor JD780A Series User Manual Download Page 292](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616292.webp)
JD780A Series Analyzers
292
Using LTE
– TDD Signal Analyzer
Figure 134
Control channel measurement with LTE - TDD signal analyzer (MBMS On)
NOTE
You can use the
LIMIT
hot key to analyze your measurements with the user-definable
limit and Pass/Fail indication. See
“Setting limit for modulation tests” on page 300 for
more information.
Using marker
You can use the
MARKER
hard key to place a marker on a resource block and display the IQ diagram
for the selected resource block.
Procedure
1.
Press the
MARKER
hard key.
2.
Press the
Channel
soft key to select the channel to be displayed in the IQ diagram.
The channel P-SS is selected by default.
3.
Turn the rotary knob to move the highlight from one to another in the channel summary table.
4.
Toggle the
Marker View
soft key between
On
and
Off
to display or dismiss the result of the
selected channel.
Subframe
The Subframe measures the modulation accuracy of all the data and control channels at the specified
subframe (1 ms).
Setting measure setup
After configuring test parameters as described in the
“Configuring test parameters” on page 265, you
can continue your measurement. The measurement settings can be saved and recalled as a file. You
can also use JDViewer, PC application software to configure a measure setup, save as a file, and load
the file on to the instrument.
Procedure
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...