![JDS Uniphase CellAdvisor JD780A Series User Manual Download Page 402](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616402.webp)
JD780A Series Analyzers
402
Using EV-DO Signal Analyzer
5.
To set the integration bandwidth for the selected index, complete the following steps:
a.
Press the
Integration Bandwidth
soft key.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Select the unit:
GHz
,
MHz
,
kHz
, or
Hz
.
6.
To add more channels to be scanned, repeat steps
3-5
.
7.
Press the
ESC
hard key to dismiss the channel list window and view the scanning result.
Setting measure setup
After configuring other test parameters as described in the
“Configuring test parameters” on page 373,
you can continue your measurement. The measurement settings can be saved and recalled as a file.
You can also use JDViewer, PC application software to configure a measure setup, save as a file, and
load the file on to the instrument.
Procedure
1.
Press the
MEASURE SETUP
hot key.
2.
To set the incremental step for the PN offset, complete the following steps:
a.
Press the
PN Increment
soft key.
b.
Enter a value by using the numeric keys.
c.
Press the
Enter
soft key.
3.
Optional
. Go to
SAVE/LOAD
> Save
, and then select the save option from the choices available
for your measurement mode. See
“Using save” on page 31 for more information.
Setting limit for OTA channel scanner
Procedure
1.
Press the
LIMIT
hot key.
2.
Press the
Limit Line
soft key to set a threshold for the limit line and Pass/Fail indication.
3.
Enter a value, and then press the
dBm
unit soft key.
4.
Toggle the
Limit Line
soft key between
On
and
Off
to display or dismiss the limit line.
5.
Optional
. You can enable alarm sound that goes off if the measurement falls outside of the limit.
Toggle the
Beep
soft key between
On
and
Off
to enable or disable the beep sound.
6.
Optional
. Go to
SAVE/LOAD
> Save
, and then select
Limit
to save the limit settings.
See
“Using save” on page 31 for more information.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...