JD780A Series Analyzers
468
Using Mobile WiMAX Signal Analyzer
emission within the necessary channel bandwidth and modulated signal measured at the RF port of
the Base Station.
Setting measure setup
After configuring test parameters as described in the
“Configuring test parameters” on page 456, you
can continue your measurement.
Procedure
1.
Press the
MEASURE SETUP
hot key.
2.
To set up the range table and parameters, press the
Range Table
soft key and then complete
the following steps:
a.
Press the
Range
soft key, and then select the range number between
1
and
20
to add
as a new or change the existing settings.
b.
Do the following as desired:
–
Select
Start Frequency
, and then specify the start frequency for the selected range.
–
Select
Stop Frequency
, and then specify the stop frequency for the selected range.
–
Select
Start Limit
, and then specify the lower limit for Pass/Fail indication.
–
Select
Stop Limit
, and then specify the upper limit for Pass/Fail indication.
–
Select
More (1/2) > Attenuation
, and then specify an value in the multiple of five.
–
Select
More (1/2) > RBW
, and then specify a RBW value.
–
Select
More (1/2) > VBW
, and then specify a VBW value.
c.
Press the
PREV
hard key.
d.
Toggle the
Range
soft key and select
On
to display the selected range in the result table or
Off
to hide it from the table.
3.
Toggle the
Measure Type
soft key between
Examine
and
Full
to select the measurement type.
NOTE
The
Examine
mode displays only the selected range while the
Full
mode lets the
instrument automatically changes the selected range from one another.
4.
To set the number of measurements to be averaged, complete the following steps:
a.
Press the
Average
soft key.
b.
Enter a value between
1
and
100
.
c.
Press the
Enter
soft key.
5.
To move the highlighted bar in the result table to other range, complete the following steps:
a.
Press the
Range
soft key in the Measure Setup menu bar.
b.
Enter a value between
1
and
20
.
c.
Press the
Enter
soft key.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...