JD780A Series Analyzers
Using CDMA Signal Analyzer
335
3.
Press the
Center Frequency
soft key.
4.
Enter a value by using the numeric keys. You can also use the rotary knob.
5.
Select the unit:
GHz
,
MHz
,
kHz
, or
Hz
.
6.
Optional
. To define the amount of frequency increment for the rotary knob, complete the following
steps:
a.
Press the
Frequency Step
soft key.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Press the unit:
GHz
,
MHz
,
kHz
, or
Hz
.
To set the channel number:
1.
Press the
FREQ/DIST
hard key.
2.
Toggle the
Unit
soft key and select
Channel
.
3.
To select the standard channel, complete the following steps:
a.
Press the
Channel Std
soft key. The standard channel window appears.
See
“Appendix C – Band, frequency & channel standard”” on page 571 for more
information.
b.
Select the band to be measured by using the rotary knob, the arrow keys, or the
Page
Up/Page Down
soft keys.
c.
Press the
Select
soft key or the rotary knob to confirm the selection.
4.
Press the
Channel Number
soft key.
5.
Enter a value by using the numeric keys. You can also use the rotary knob.
6.
Press the
Enter
soft key.
The instrument automatically displays the corresponding center frequency value for the selected
channel number.
7.
Optional
. To define the amount of channel increment for the rotary knob, complete the following
steps:
a.
Press the
Channel Step
soft key.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Press the
Enter
soft key.
NOTE
This frequency setting is not used in the Multi-ACLR and Spurious Emissions modes.
Setting amplitude
Reference level and attenuation
You can set the reference and attenuation levels automatically or manually to optimize the display of
the traces measured, as you desire.
Procedure
To automatically set the reference and attenuation level:
1.
Press the
AMP/SCALE
hard key.
2.
Press the
Auto Scale
soft key.
Each time you press this key, both of the Y-axis scale and input attenuation level change to be
optimized with some margin.
To set the reference or attenuation level manually:
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...