431
Power vs. time (frame)
The Power vs. Time (Frame) measures the modulation envelope in the time domain, showing the
power of each time slot of the TD-SCDMA signal.
Setting measure setup
After configuring test parameters as described in the
“Configuring test parameters” on page 415, you
can continue your measurement. The measurement settings can be saved and recalled as a file.
Procedure
1.
Press the
MEASURE SETUP
hot key.
2.
Press the
Slot Number
soft key, and then select the time slot number to be displayed:
TS0
,
TS1
,
TS2
,
TS3
,
TS4
,
TS5
, or
TS6
.
3.
Press the
Spread Factor
soft key, and then select the spread factor option:
DL Auto
for auto
detection of downlink,
UL Auto
for auto detection of uplink,
SF1
,
SF2
,
SF4
,
SF8
, or
SF16
4.
To set the number of maximum users, complete the following steps:
a.
Press the
Max Users
soft key.
b.
Enter a value up to 16 by using the numeric keys.
c.
Press the
Enter
soft key.
5.
Optional
. To set the delay, complete the following steps:
a.
Press the
Delay
soft key to set the amount of delay in µs.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Press the
µs
soft key.
NOTE
The Delay setting is used only when there is a time offset in the signals to be measured.
6.
Optional
. Press the
Miscellaneous
soft key, and then do the following(s) as needed:
–
To select how to determine scrambling estimation, toggle the
Scramble
soft key and select
Spec
or
Full
.
NOTE
The Spec searches
“Sync Sequence Number” from 0 to 31 based on the TD-SCDMA
specification and determines
“Scrambling Code Number” by using the sync sequence
number, while the Full does from 0 to 127 fully.
–
To select how to determine the Midamble spread factor estimation either in full or maximum
power search, toggle the
Mid SF Map
soft key and select
Max
or
Spec
.
NOTE
The Max determines the spread factor of the Midamble by using the maximum power of
the received Midamble, while the Spec searches
“Default Midamble Allocation” based on
the TD-SCDMA specification.
–
To select the detect mode, press the
Detect Mode
soft key and then select the option:
Auto
,
QPSK/8PSK
,
QPSK
,
8PSK
,
16 QAM
, or
64 QAM
.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...