![JDS Uniphase CellAdvisor JD780A Series User Manual Download Page 305](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616305.webp)
JD780A Series Analyzers
Using LTE
– TDD Signal Analyzer
305
This MIMO (Multiple Input Multiple Output) setting is activated only if the option 031 LTE-
Advanced TDD is installed with a license number. If not, the instrument sets this option to
2x2 by default. A 2x1 or 4x1 RF combiner is also required to able to test on MIMO
channels.
d.
Press the
Antenna Port
soft key, and then select the option:
Auto
,
0
,
1
,
2
, or
3
.
If the option 031 LTE-Advanced TDD is not installed or if the MIMO is set to 2x2, the
antenna ports 2 and 3 are disabled.
e.
Toggle the
External Offset
soft key and select
On
.
f.
Enter a value by using the numeric keys, and then press the
dB
soft key.
g.
To turn the Multimedia Broadcast Multicast Service (MBMS) feature on or off, toggle the
MBMS
soft key and select
On
or
Off
. For proper MBMS testing, you need to set the
Subframe No with the same PMCH subframe number.
h.
To set the Multicast Broadcast Single Frequency Network (MBSFN) detection automatically
or manually for MBMS testing, toggle the
MBSFN
soft key and select
Auto
or
Manual
. An
automatically detected or manually entered MBSFN ID appears on the screen.
8.
To save your settings and results, go to
SAVE/LOAD
> Save
and then perform functions as you
desire. See
“Using save” on page 31 for more information.
9.
Press the
Run Test
soft key to start to run a test.
The Auto Measure Results window appears at the end of the test.
10.
To stop running the test, press the
Abort
soft key.
11.
To change the view on the screen during the test, press the
Display
and then select the view
option from the following choices:
–
Screen
: You can view each measurement screen as the test progresses.
–
Results
: You can view a measurement result table as the test progresses.
–
Settings
: You can view a measurement setting table as the test progresses.
Setting display
After completion of the auto measurement, the screen menu changes to Trace/Display so that you can
view the results in different forms.
Procedure
1.
Toggle the
Display
soft key and select the display option:
–
Result
:
to view the result table. The Display Result menu becomes activated.
–
Settings
:
to view the measurement settings for the auto measurement.
2.
Toggle the
Display Result
soft key and select the display result option:
–
Full
: to view detailed measurement readings with the pass/fail indication.
–
Quick
: to view only the Pass/Fail results.
–
Compare
: to compare two measurement values for each test time with the Allowed %
Change setting on. You can view results vs. factory-defined or user-defined references.
3.
To view the measurement results for a different carrier, press the
View Carrier
soft key and then
select the carrier number to view.
4.
Press the
Replace Reference
soft key to replace existing reference values for comparison with
new measurement data.
5.
Press the
Load From
soft key to select the location from which references file to be loaded.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...