![JDS Uniphase CellAdvisor JD780A Series User Manual Download Page 218](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616218.webp)
JD780A Series Analyzers
218
Using LTE - FDD Signal Analyzer
NOTE
You can use the
LIMIT
hot key to analyze your measurements with the user-definable
limit and Pass/Fail indication. See
“Setting limit for modulation tests” on page 231 for
more information.
Using marker
You can use the
MARKER
hard key to place a marker on a resource block and display the IQ diagram
for the selected resource block.
Procedure
1.
Press the
MARKER
hard key.
2.
Press the
RB
Number
soft key to select the resource block to be marked.
3.
Enter the resource block number by using the numeric keys.
4.
Press the
Enter
soft key.
The marker appears on the selected resource block.
5.
Toggle the
Marker View
soft key between
On
and
Off
to display or dismiss the result of the
selected resource block.
Control channel
The Control Channel measures the constellation for the specified control channel as well as
modulation accuracy of the control channel at the specified subframe.
Setting measure setup
After configuring test parameters as described in the
“Configuring test parameters” on page 197, you
can continue your measurement. The measurement settings can be saved and recalled as a file.
Procedure
1.
Press the
MEASURE SETUP
hot key.
2.
To set the nominal channel bandwidth to be measured, complete the following steps:
a.
Press the
Bandwidth
soft key.
b.
Select the bandwidth option:
1.4 MHz
,
3 MHz
,
5 MHz
,
10 MHz
,
15 MHz
, or
20 MHz
.
3.
To select the subframe number to be measured, complete the following steps:
a.
Press the
Subframe No
soft key.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
NOTE
For MBMS testing, this subframe number must be neither 0 nor 5 as these subframes 0
and 5 are not available for MBMS and it must be set to the MBMS transmitted channel
’s
subframe number.
c.
Press the
Enter
soft key.
4.
Optional
. Toggle the
Cell ID
soft key and select
Auto
or
Manual
as desired:
–
Auto
: to let the instrument detect the Cell ID for the LTE signal automatically.
–
Manual
: to set a specific Cell ID for the LTE signal manually in order to speed up the
synchronization with a BTS.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...