![JDS Uniphase CellAdvisor JD780A Series User Manual Download Page 337](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616337.webp)
JD780A Series Analyzers
Using CDMA Signal Analyzer
337
Pre-amplifier
You can turn the internal pre-amplifier on to correct and compensate for the gain of the preamp so that
amplitude readings show the value at the input connector.
Procedure
1.
Press the
AMP/SCALE
hard key.
2.
Toggle the
Preamp
soft key and select
On
or
Off
as needed.
NOTE
You can turn the Preamp on when the input attenuation range is from 0 dB to 10 dB. If
the attenuation value is manually set to greater than 10 dB, the instrument will
automatically turn off the pre-amplifier to display low-level signal properly on the chart.
External offset
You can turn the
External Offset
on and manually set the external offset value. An offset consists of a
cable loss and a user offset and the measurement result shows the value reflecting both offset values.
When the external offset value is set at 40 dB in the Spectrum mode, the measurement result
compensates 40 dB at both the Spectrum Analyzer and Signal Analyzer modes
Procedure
To set the external offset:
1.
Press the
AMP/SCALE
hard key.
2.
Toggle the
External Offset
soft key and select
On
.
3.
Enter a value by using the numeric keys.
4.
Press the
dB
soft key to complete the entry.
To turn the external offset off:
1.
Press the
AMP/SCALE
hard key.
2.
Toggle the
External Offset
soft key and select
Off
.
Setting average
You can set the number of measurements to be averaged for the trace presentation. A maximum of
100 times of averaging can be set. When the averaging reaches to your setting, a new measurement
value replaces the measurement value in sequence from the earliest.
Procedure
1.
Press the
BW/AVG
hard key.
2.
Press the
Average
soft key.
3.
Enter a value between 1 and 100 as needed by using the numeric keys.
4.
Press the
Enter
soft key.
Setting sweep mode
The default setting is
Continue
to sweep continuously for most on-going measurements. If you want to
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...