OB Report Page
190
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Osaka University Report Page (cont’d)
Exam date, ID, Name, Referral, Referred For, Age,
Gravida and Para: These fields are identical to the US
report page as described on page 176.
Pregnancy Origin: The pregnancy origin field is similar
to Tokyo University as described on page 185.
EFBW: The Estimated Fetal Body Weight is computed
using two distance calculations (BPD & FL) and one area
calculation (FTA).
Estimated Fetal Body Weight is automatically computed if
the above measurements have been made. Calculations
are made from the last measurement entered or averaged
using the formula:
EFBW[g] = 1.25647 x BPD[cm]
3
+ 3.50655 x FTA[cm
2
]
x FL[cm] + 6.3
GA (EFBW): The Gestational Age by Estimated Fetal
Body Weight is automatically computed if measurements
for EFBW are already made (BPD, FL and FTA). Please
refer to the EFBW (Osaka) table on page 342 for details.
EFBW (IU): If the standard deviation value for EFBW is
greater than 1.5, the EFBW (IU) is computed using the
formula: ;
EFBW(IUGR)[g] = 1.229 x BPD[cm]
3
+ 3.063 x FTA[cm
2
]
x FL[cm] – 24.6
GA (EFBW IU): Based on the calculations made for
EFBW(IU), the GA by EFBW (IU) is automatically
computed. Please refer to the EFBW (Osaka) table on
page 342 for details.
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...