Cardiac Measurements
230
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
ta
Substitution of a LV measurement
A previous LV measurement like LVIDd, LVIDs or the
Heart Rate can be replaced independently without
repeating the whole sequence.
Example: After completion of the Cubed formula LV functions it
is realised that LVIDs has been erroneously measured.
To replace LVIDs without repeating the whole sequence:
Press [4]. The LVIDs item is highlighted. An open ended “ ”
cursor with a vertical line appears on the screen.
Use the TRACKBALL to move the open ended “ ” cursor to
either end of the starting point of LVIDs dimension.
Press SET to fix the start point cursor. A second open ended
“ ” cursor appears on the vertical line. It is moveable only in
the vertical axis.
Use the TRACKBALL to move the open ended “ ” cursor to
the end point of the LVIDs measurement.
Press SET to fix the end point cursor. The LVIDs is corrected
and the corresponding ESV, SV, EF, and CO are displayed.
TB
TB
4
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...