Biopsy Procedures
288
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Needle Guide Type Preset
Activate the biopsy guidelines on the system through
CONTROL B as explained earlier. Press CONTROL N or
the alphanumeric key ‘N’ (user should not be in Com-
ment mode) to choose the type of biopsy guide angle
(guide zone) display to conform with desired target
depth/location. Selection for respective probes is
available from the chart below.
Depth in cm at Center Channel
Fixed Angle
Multi–Angle
Probe
SGL
MBX–1
MBX–2
MBX–3
C36
8.0
4.0
6.0
8.0
C55
7.0
4.0
5.5
7.0
L76
2.0
2.0
4.0
7.0
C31
6.0
n/a
n/a
n/a
VE5
n/a
n/a
n/a
n/a
Table 41. Biopsy Guide Attachment Selection
Depth in cm at Center Channel
Multi–Angle
Probe
TV0
°
TR5
°
E72
10.0
13.3
Table 42. Biopsy Guide Attachment Selection
Failure to match the preset to the guide may cause the
needle to track a path outside the displayed lines.
It is extremely important that when using the adjustable
angle biopsy guides, the preset displayed on the screen
matches the angle set on the guide, otherwise the
needle will not follow the displayed guide zone which
could result in repeated biopsies or patient injury.
DANGER
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...