Basic Measurements
113
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Measurement of Volume by approximation to a Sphere (One
Distance Method)
This method is appropriate for structures having radial
symmetry.
Pre Selection
Scan and obtain a cross section of the structure the
volume of which is to measured.
Measure the diameter. (For details of distance
measurement please refer page 103).
Press 2
The volume is calculated and displayed on the left side of
the screen in the calculation result area. The
measurement d
1
used in the calculation is also displayed.
Post Selection
Scan and obtain a cross section of the structure the
volume of which is to measured.
Press
2. The
2:VOL
item gets highlighted.
Measure the diameter. (For details of distance
measurement please refer page 103).
Press
CLEAR. The volume is calculated and displayed
on the left side of the screen in the calculation result
area. The measurement d
1
used in the calculation is also
displayed.
2
2
2:VOL
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...