Biopsy Procedures
287
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Biopsy Procedure (cont’d)
NOTE: Press the MEASUREMENT key once while the
guidelines are present to display the integrated biopsy
depth cursor and center line. Use the TRACKBALL to
measure the needle length. Needle length is from the top
of needle barrel to the target.
2
1
1
Needle Length
2
Target
Illustration 56
The needle may vary from the center line or guidezone
for various reasons:
S
Needle barrel to needle clearance or strength.
S
Bracket manufacturing tolerance.
S
Needle deflection due to tissue resistance.
S
Needle size chosen. Thinner needles may deflect more.
S
Wrong needle guide type selection (using Ctrl N key).
The display should be carefully monitored during a
biopsy for any needle deviation from the center line or
guidezone.
NOTE: The acceptable needle deviation is between
2 mm and 4 mm. If the deviation is greater than this, call
Service.
.
CAUTION
.
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...