HIP Dysplasia (HIP)
The HIP calculation assists in assessing the
development of the infant hip. In this calculation three
straight lines are superimposed on the image and
aligned with the anatomical features. The two angles are
computed, displayed and can be used by the physician
in making a diagnosis.
The three lines are (Source: R GRAF, journal of Pediatric
Orthopedics, 4: 735–740 (1984)).
S
The inclination line connects the osseous convexity
to the labrum acetabulare.
S
The Acetabulum roof line connects the lower edge of
the osileum to the osseous convexity.
S
The Baseline connects the osseous acetabulum
convexity to the point where the joint capsule and the
perichondrium unite with the iliac bone.
The Angle
T
(Alpha) is the supplement of the angle
between 1 and 3. It characterizes the osseous convexity.
The angle
b
(Beta) is the angle between lines 1 and 2. It
characterizes the bone supplementing additional roofing
by the cartilaginous convexity.
b
T
Illustration 33. HIP Dysplasia
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...