Troubleshooting
Overview
Listed in this section are problems or system messages
that may be encountered, possible causes for the
problem or message, and the appropriate action to take
to correct the situation. Consult
Who to Contact on page
313
if additional information or assistance is needed.
Troubleshooting the LOGIQ
α
100
Problem
Cause
Solution
When Power is turned ON,
the screen still appears
bl
k
Power Cable may have
come out from socket
Reconnect the Power
Cable.
blank.
The Monitor brightness and
contrast may have been
tampered with.
Adjust the Monitor
brightness and contrast.
The power fuse may have
blown.
Turn off power switch.
Replace power supply unit.
No image appears on the
screen.
The
FREEZE key may be
on.
Check the
FREEZE key
and turn it off.
Dark vertical stripes appear
on the image
The probe may be cracked
or particles may be stuck
to the surface of the probe.
Clean the surface of the
probe.
Blinking dots of noise
appear on the image.
The
LOGIQ
α
100
may
be placed near other
ultrasonic units.
Move the
LOGIQ
α
100
away from other units.
Wavering image on the
display screen.
LOGIQ
α
100
may be
close to some magnetic
field.
Move the unit away from
any magnetic field.
No scanned image
appears when probe is
placed on a patient.
The Gain control may be at
0dB.
Increase the Gain.
Table 45. Console Troubleshooting
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...