Probes
264
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Probe Cleaning Process (
cont’d
)
3. Rinse the probe with enough clean potable water
to remove all visible soap residue.
4. Air dry or dry with a soft cloth.
Special Cleaning Instructions for the E72: When
cleaning the E72 probe, it is important to be sure that all
surfaces are thoroughly cleaned. This probe has an
adjustable two-part handle that must be disassembled to
gain access to all surfaces. To disassemble the handle,
completely remove the handle adjustment screw located
mid-way between the cable entry and probe tip the two
handle halves and adjustment screw must be thoroughly
cleaned along with the main probe shaft as described
earlier in step 2. After rinsing and drying is completed,
the probe handle can be loosely reassembled for the
disinfection process.
C36
L76
E72
Aperature
Fluid Level
Contact face within
Patient Environment
C55
C31
VE5
Illustration 54. Probe Immersion Levels
.
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...