Cardiac Measurements
229
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Measurement of LV Functions (cont’d)
Use the Trackball to move the second cursor to the point
of diastole on the diametrically opposite wall (inter
ventricular septum).
Press Set. The measurement is completed and the
dimension appears besides the LVIDd item.
The EDV is calculated and displayed simultaneously
besides the EDV item.
Simultaneously a third cursor appears and the
highlighted item on the list shifts to LVIDs.
Repeat measurements as above.
On completion the measurement is recorded in LVIDs
and the calculated ESV, SV and EF are displayed along
side their respective items.
A fifth cursor appears and the highlighted item in the
menu is heart rate (HR).
Measurement of heart rate is by the same procedure as
mentioned in heart rate measurement (page 118).
On completion of the heart rate measurement, the
cardiac output (CO) is displayed.
Toggling between cursors is not possible on pressing
MEASURE.
To clear a measurement and apply a second formula,
press
CLEAR repeatedly until the measurements are
erased. Then start a new session.
NOTE: To abort a measurement midway press CLEAR.
Unfreezing the screen midway clears the previous LV
function measurement.
TB
Repeat for LVIDs
and heart rate
Hints
.
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...