Cardiac Measurements
224
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
ta
Amplitude measurement
The default measurement in M
–
Mode is amplitude.
The amplitude of a moving cardiac structure is the
greatest distance it has moved in a well defined phase of
the cardiac cycle and is expressed in millimeters.
Amplitude measurements are done only in M
–
Mode.
Perform a scan in B/M or M
–
Mode.
Press
FREEZE to stop image acquisition.
Press the MEASUREMENT key once to display the
measurement menu and an open ended “ ” cursor
appears on the screen.
Use the
TRACKBALL to move the open ended “ ” cursor
to the amplitude measurement start point.
Press set to fix the measurement start point. A second
cursor appears. Amplitude in mm appears on the left
side of the screen.
Use the
TRACKBALL to move the second open ended
“ ” cursor to the measurement end point.
Press
SET to complete the measurement. The amplitude
value is displayed in mm on the left of the screen.
Repeat the same procedure to make more amplitude
measurements, if required.
Up to 4 measurements will
be displayed in mm at a time.
+
+
amplitude
M
A
"
Meas
TB
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...