Probes
267
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Coupling gels
Applying
In order to assure optimal transmission of energy
between the patient and probe, a conductive gel or
couplant must be applied liberally to the patient where
scanning will be performed.
Precautions
Coupling gels should not contain the following
ingredients as they are known to cause probe damage:
S
Methanol, ethanol, isopropanol, or any other
alcohol-based product
S
Mineral oil
S
Iodine
S
Lotions
S
Lanolin
S
Aloe Vera
S
Methyl or Ethyl Parabens (para hydroxybenzoic acid)
Planned Maintenance
The following maintenance schedule is suggested for the
system and probes to ensure optimum operation and
safety.
Do the
Following
Daily
After
Each Use
As Necessary
Inspect the
Probes
X
Clean the Probes
X
Disinfect Probes
X
Table 21. Planned Maintenance Program
CAUTION
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...