INSTALLATION SETUP
1
.
HOSPITAL NAME
:
2
.
DATE FORMAT: 1
1:DD/MM/YY 2:MM/DD/YY 3:YY/MM/DD)
3
.
DATE: 4.TIME
:
5.OB VERSION SELECTED: 1
(1:US 2:TOKYO 3:OSAKA 4:EUROPE 5:ASUM)
6.FILM EXPOSURE TIME: 4
(1:125ms 2:250ms 3:375ms 4:500ms)
7.MINIMUM FILM EXPOSURE INTERVAL : 2 (1–9 seconds)
8.VIDEO INVERT FOR REPORT PRINT : 2 (1:YES2:NO)
9.CIRCUMFERENCE MEASUREMENT METHOD: 1
(1:ELLIPSE 2:2DISTANCE 3:TRACE)
XX/XX/XX
XX:XX
10. US GA SELECTION : 1(1:CUA 2:AUA)
11. ADD 1 WEEK TO EDD : 1 (1:NO 2:YES)
12. LANGUAGE : 1 (1:ENG 2:GER 3:FRE 4:ITA 5:POR 6:SPA)
13. HIP ORIENTATION:1 (1:CRANIAL LEFT 2:CAUDAL LEFT)
Illustration 51 Installation Setup Menu
EUROPEAN OB TABLE SETUP
Item:Sel
Option
Author
GS:
CRL:
BPD:
HC:
AC:
FL:
BD:
TAD:
OFD:
Ft:
EFW:
9
1
1
1
1
1
1
3
5
3
1
(0,2,9)
(0,1,2,3,4,6)
(0,1,2,3,4,5,7)
(0,1,2,5)
(0,1,2,5)
(0,1,2,3,4,5)
(0,1,4,5,8)
(0,2,3)
(0,2,5)
(0,3)
(1:SHEP/WARS 2:RICH/BERK 3:HADLOCK 4:GERMAN)
1: JEANTY
2: HANSMANN
3: PARIS
4: CAMPBEL
5: SOSTOA
6: NELSON
7: KURTZ
8. BERKOWITZ
9. TOKYO
0: N/A
Illustration 52 European Author Selection
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...