OB Report Page
186
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Tokyo University Report Page (cont’d)
EFBW: Estimated Fetal Body Weight is automatically
computed if BPD, APTD, TTD and FL measurements are
made. If any of the above measurements are not
available, the EFBW field will be blank. EFBW is
calculated using the formula:
EFBW = 1.07xBPD^3+3.42xAPTDxTTDxFL
GA (Pregnancy Origin)
Depending on the parameter chosen for pregnancy origin
the Gestational Age is calculated as described below.
GA (LMP): If LMP is selected in Pregnancy Origin, the
field GA(LMP) appears. The Gestational Age by Last
Menstrual Period is computed from the date of the first
day of the Last Menstrual Period, according to the
formula:
GA (LMP) = {Exam date} – {Last menstrual period}
GA (EDC): If EDC is selected in Pregnancy Origin, the
field GA(EDC) appears. The Gestational Age by
Expected Date of Confinement is computed from the
formula:
GA (EDC) = {EDC} + {Exam date} – {40 weeks or 41
weeks as selected in setup}
GA (BBT): The Gestational Age by Basal Body
Temperature is computed using the formula:
GA (BBT) = {Exam date} – {BBT} + 2 weeks
CGA: The Composite Ultrasound Age is calculated using
the formula
CGA
+
G
1
) G
2
) G
3
) ... ) G
n
n
where G
1
, G
2
, ... G
n
are the Gestational Ages obtained
from the corresponding measurements selected for use
in CGA calculation and n is the total number of such
measurements used.
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...