Disinfecting probes
Ultrasound probes can be disinfected using liquid
chemical germicides. The level of disinfection is directly
related to the duration of contact with the germicide.
Increased contact time produces a higher level of
disinfection. 2% Glutaraldehyde-based solutions have
been shown to be very effective for this purpose. Cidex is
the only germicide that has been evaluated compatibility
with the material use to construct the probes.
In order for liquid chemical germicides to be effective, all
visible residue must be removed during the cleaning
process. Thoroughly clean the probe, as described
earlier, before attempting disinfection.
1. Prepare the germicide solution according to the
manufacturer’s instructions. Be sure to follow all
precautions for storage, use and disposal.
2. Place the cleaned and dried probe surface in
contact with the germicide solution. High-level
disinfection (for surface and cavity probes) can be
accomplished by soaking according to the contact
time recommended by the germicide
manufacturer.
3. After removing from the germicide, rinse the
probe following the germicide manufacturer’s
rinsing instructions. Flush germicide residue from
the probe and allow to air dry.
After Each Use
CAUTION
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...