OB Measurements
156
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Fetal Trunk Cross Sectional Area (FTA)
Fetal Age Estimation is arrived at by measuring the
cross–sectional area of the fetal trunk. Two measurement
procedures (pre
–
selection and post
–
selection) help
achieve flexibility in making FTA measurements.
Pre
–
Selection: The user selects the FTA measurement
item first and then performs the measurement. The FTA
measurement can be arrived at by using Two Distance,
Ellipse or Trace Method.
S
If FTA (AC and HC) are to always be measured using
the Two Distance method, select the Circumference
Measurement Method in the Installation Setup as
Two Distance. Then FTA selection will only use the
Two Distance Method.
S
If FTA (AC and HC) are to always be measured by
the Ellipse Method, select the Circumference
Measurement Method in the Installation Setup as
Ellipse. Then FTA selection (AC and HC) will only use
Ellipse Method.
S
Similarly to measure FTA (AC and HC) by the Trace
Method, select Trace in the Installation Setup.
For details on Installation Setup, refer to the System
Setup section on page 33.
+
+
3
Take
Measurement
Summary of Contents for LOGIQ A100
Page 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Page 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Page 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Page 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Page 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Page 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Page 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Page 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Page 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Page 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...