Methods 5
Notes:
Methods that include the
InjectAndRecover
command require a sample
and reagent rack and cannot be used with microplates (see the
Biacore T100 Instrument Handbook for rack details).
The contact time for sample, flow rate and number of repetitions
determine the total injected volume for both sample and recovery
solution. You may need to adjust one or more of these parameters if the
method does not pass verification.
Regeneration command
This command is intended for injection of a regeneration solution following the
sample injection. Check
High Viscosity Solution
if the regeneration solution has
a relative viscosity higher than about 3 (corresponding to about 35% glycerol or
40% ethylene glycol at 20°C). This will adapt the solution aspiration and injection
procedure for higher viscosity. The injected solution, contact time and flow rate
can be set as variables.
Sample command
This command is intended for injection of sample containing analyte. Only
Sample
commands are recognized as analyte injections in the Evaluation
Software for kinetics, affinity and concentration evaluation. The injected
solution, contact time, dissociation time and flow rate can be set as variables.
Evaluation variables can also be defined for the
Sample
command (see
Section 5.6.2).
The
Sample
command offers three alternative settings under
Type
:
High performance
Optimizes the injection for high performance by using
extra segments of air and sample during aspiration to
separate the injected solution from running buffer,
thereby minimizing dispersion of sample at the
beginning and end of the injection at the expense of
additional 25 µl sample consumption.
Low sample
consumption
Optimizes the injection for low sample consumption by
using fewer air segments than the High performance
setting. The Low sample consumption setting however
still achieves a performance that is adequate for most
applications except analysis of rapid kinetics.
Summary of Contents for Biacore T100
Page 1: ...GE Healthcare Biacore T100 Software Handbook ...
Page 2: ......
Page 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Page 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Page 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Page 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 237: ......