Biacore T100 Software Handbook BR-1006-48 Edition AE
81
Methods 5
Concentration unit
This setting defines the unit for entering sample concentrations. The unit must
be specified here, and cannot be changed at any other step in the assay
definition. The unit can however be changed in the Evaluation Software when
the results of the run are evaluated.
Buffer settings
Enter names if desired for the buffers in bottles A to D. Names entered here will
be displayed in the
Prepare Run Protocol
(Section 4.2.7). Different buffers may
be chosen for different assay steps (Section 5.5.4).
Specify analysis temperature after run
Check this option and enter a temperature to set the analysis temperature
when the run is completed. The rack temperature will also be reset if the
Vary
with analysis temperature
box is checked. This setting provides automated
control of the chip and detector environment after the completion of a run, for
example in preparation for another run at a different temperature.
5.5
Assay steps
This screen determines the main structure of the method in terms of assay
steps. Steps at the top level (i.e. not indented or marked with the symbol ) are
executed in the order given. Nested steps (marked with the symbol ) are
executed within the context of the level in which they are placed, as specified by
the settings for
Recurrence
.
To create a new assay step, click
New Step
. The step will be created with default
settings at the end of the current method. Move the step in the method with the
Summary of Contents for Biacore T100
Page 1: ...GE Healthcare Biacore T100 Software Handbook ...
Page 2: ......
Page 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Page 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Page 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Page 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 237: ......