4 Application wizards
4.3 Surface preparation wizards
50
Biacore T100 Software Handbook BR-1006-48 Edition AE
Select an injection and use the
Edit
,
Delete
,
Move up
and
Move down
buttons
to edit the injection details, remove the injection from the method and change
the order of injections in the method.
Custom methods are stored in the immobilization wizard template: if you need
the same or slightly modified method in a different template, save a copy of the
template and then edit the method.
Step 2. System preparations
Check the
System preparations
options as required (see Section 4.2.5).
Step 3. Rack positions
Accept or change the rack positions for the various solutions required (see
Section 4.2.6).
Pre-conc
injects 10 µl of ligand solution at 5 µl/min to estimate the rate of
preconcentration. This step is only performed if the option
Aim
for immobilized level
is chosen when the immobilization method
is used. A method may only contain one
Pre-conc
injection. The
Pre-conc
injection should always be placed before surface
activation: it will usually be first in the method, although it may be
preceded by a surface conditioning injection if required. If you
place the
Pre-conc
injection after the surface activation, it will be
executed there and the ligand will be immobilized on the
activated surface.
After the
Pre-conc
injection, the surface is washed with a solution
specified in the immobilization setup dialog, to remove any ligand
that may remain on the surface.
Do not use a
Pre-conc
injection with Sensor Chip SA, since
biotinylated ligand will bind to the surface and cannot be
removed.
Inject
performs an injection of a specified solution with a specified
contact time and flow rate. Values are entered in the dialog box
that appears when you click
Inject.
Mix & Inject
mixes two specified solutions and performs an injection of the
mixture. Details are entered in the dialog box that appears when
you click
Mix & Inject.
Wash
washes the flow system (but not the sensor surface). The wash
solution is specified in the dialog box that appears when you click
Wash.
Summary of Contents for Biacore T100
Page 1: ...GE Healthcare Biacore T100 Software Handbook ...
Page 2: ......
Page 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Page 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Page 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Page 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 237: ......