Biacore T100 Software Handbook BR-1006-48 Edition AE
87
Methods 5
General command properties
Common features of several commands are check-boxes for
Predip
,
Extra
Wash After Injection
and
Stabilization period.
Capture command
This command is intended for injection of ligand over a capturing molecule at
the beginning of a cycle. The injected solution, contact time and flow rate can
be set as variables.
Carry-over control command
This command injects a 30-second pulse of buffer over the surface at a flow rate
of 40 µl/min, in order to check that there is no carry-over of analyte or other
material from an injection earlier in the cycle. The injection is suitably placed at
the end of the cycle, and can be used in a conditional context (see the
If…then
command) to perform additional buffer injections or regeneration steps if carry-
over is detected. A plot of the binding response from a carry-over injection
against cycle number is created automatically for quality control purposes in
the evaluation software (see Section 6.4).
If…then command
This command allows construction of conditional methods, where commands
are executed or skipped depending on the outcome of certain conditions. The
illustration below shows a cycle which will perform an additional regeneration if
the relative response after the first regeneration exceeds a specified value:
Predip
Check this box to dip the needle in a separate position before
aspirating the solution to be injected. The predip position will
normally contain the same solution as is injected, so that the
needle is rinsed briefly to minimize carry-over effects. The
same predip position is used for all cycles.
Extra Wash
After Injection
Check this box and specify a wash solution to perform an
extra wash of the flow system after the injection. The flow
system is washed automatically with buffer after each
injection, but an extra wash with a different solution can be
included if required. This wash solution does not pass over the
sensor surface.
Stabilization
period
Check this box and specify a time in seconds to introduce a
delay before the next command is started. This can sometimes
be necessary (for example after regeneration steps) to allow
the response to stabilize before performing the next injection.
Summary of Contents for Biacore T100
Page 1: ...GE Healthcare Biacore T100 Software Handbook ...
Page 2: ......
Page 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Page 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Page 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Page 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 237: ......