Biacore T100 Software Handbook BR-1006-48 Edition AE
85
Methods 5
5.5.4
Assay step preparations
Temperature
This value determines the analysis temperature for the assay step. The setting
will also control the sample compartment temperature if the appropriate option
is checked under
General Settings
(Section 5.4).
If the actual temperature at the start of an assay step does not match the
setting for the step, the system will wait until the setting is reached.
Buffer
Select the running buffer to be used for the assay step. The default buffer is A
(corresponding to buffer bottle and tubing A on the instrument).
5.6
Cycle types
Cycle types define the detailed sequence of operations to be performed in each
assay step.
Top level step
Number of cycles
10
20
10
Number of replicates
1
1
2
Total number of cycles
10
20
20
Number of recurrences for the nested step
2
4
4
Summary of Contents for Biacore T100
Page 1: ...GE Healthcare Biacore T100 Software Handbook ...
Page 2: ......
Page 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Page 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Page 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Page 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 237: ......